Syringe pump for enteral nutrition

Code produit: SN-50C6

This syringe-driver is dedicated to the administration of enteral nutrition, only for use with syringes for enteral use.

The syringe-driver is pre-calibrated with the syringes of brands C-GON (channel 2),  A-VY (channel 3), VY-P (channel 4) and Nutrifit (channel 5). Additional channels can be used to calibrate other brands of syringes, by following the procedure detailed in the instructions for use (dedicated to biomedical engineers).

The syringe-driver can be used with following volume syringes: 10 ml, 20 ml, 30 / 35 ml and 50 ml.

Display range of volume delivered: 0.1 to 9 999 ml. 

Occlusion detection pressure : 
  • High (H) : 800 mmHg +/- 200 mmHg 
  • Center (C) : 500 mmHg +/- 100 mmHg 
  • Low (L) : 300 mmHg +/- 100 mmHg 
Power voltage : AC 110 V - 230 V 

Battery voltage: DC 12 V - A fully charged battery allows to use the syringe-driver for more than 6 hours at a speed of 5 ml/hour. 

The battery must be charged for minimum 4 hours. 

Weight : 2.3 kg 


  Code Channel 2 / CGON 10ml Channel 2 / CGON 20ml Channel 2 / CGON 30/35ml Channel 2 / CGON 50ml Channel 3 / AVY 10ml Channel 3 / AVY 20ml Channel 3 / AVY 50ml Channel 4 / VYP 10ml Channel 4 / VYP 20ml Channel 4 / VYP 50ml Channel 5 / Nutrifit 10ml Channel 5 / Nutrifit 20ml Channel 5 / Nutrifit 50ml
SN-50C6 1015.102 1015.202 1015.352 1015.602 1015.103 1015.213 1015.603 1015.104 1015.204 1015.604 1015.107 1015.207 1015.607
  Channel 2 / CGON Channel 3 / AVY Channel 4 / VYP Channel 5 / Nutrifit
Code 10ml 20ml 30/35ml 50ml 10ml 20ml 50ml 10ml 20ml 50ml 10ml 20ml 50ml
SN-50C6 1015.102 1015.202 1015.352 1015.602 1015.103 1015.213 1015.603 1015.104 1015.204 1015.604 1015.107 1015.207 1015.607
  • Contains Latex: No
  • Contains DEHP: No
  • Contains biogical or animal-based product: No