Accueil»Catalogue»I.V. Accessories»Extension tube (multi-lumen)»Double lumen extension tube with Bionector
Protect-a-set / Octopus 2
Protect-a-set (code 836.02) and Octopus 2 (other codes) are polyurethane, transparent, supple with low dead space double-lumen extension tubes for IV catheters (peripheral or central venous) fitted with one or 2 Bionector.
To identify the different infusion lines, each line has a hub or a clamp of a different colour.
They allow to connect to the catheter several independant IV lines.
Code 836.02 is a double lumen extension tube with 1 non-return valve, 1 hub with Bionector and 1 clamp (white).
Length of tubes: 6 cm / 10 cm.
Code 841.232 is a double lumen extension tube with 2 non-return valves, 2 Bionector and 2 clamps (green).
Length of tubes: 3 cm / 3 cm.
Code 841.264 is a double lumen extension tube with 2 Bionector and 2 clamps (green and orange).
Length of tubes: 10 cm / 10 cm.
Code 842.209 is a double lumen extension tube with 1 non-return valve, 2 Bionector and 2 clamps (green and orange). The end without non-return valve allows the sampling blood.
Length of tubes: 6 cm / 6 cm
Used for peripheral cannula care, Octopus 2 has the following advantages:
Non-return valves avoid reflux of fluids or drugs beyond the valve during multiple infusions.
Bionector can be disinfected with most of the antiseptic solutions (example: 70% alcohol). Avoid using antiseptic containing both alcohol and hydrogen peroxide.
For more information about Bionector, please refer to technical file code 896.
Octopus with Bionector are also available with 1 lumen (code 5222.014), 3 lumen (codes 841.364-842.311/312), 4 lumen (code 842.414) and 5 lumen (code 842.514).
To identify the different infusion lines, each line has a hub or a clamp of a different colour.
They allow to connect to the catheter several independant IV lines.
Code 836.02 is a double lumen extension tube with 1 non-return valve, 1 hub with Bionector and 1 clamp (white).
Length of tubes: 6 cm / 10 cm.
Code 841.232 is a double lumen extension tube with 2 non-return valves, 2 Bionector and 2 clamps (green).
Length of tubes: 3 cm / 3 cm.
Code 841.264 is a double lumen extension tube with 2 Bionector and 2 clamps (green and orange).
Length of tubes: 10 cm / 10 cm.
Code 842.209 is a double lumen extension tube with 1 non-return valve, 2 Bionector and 2 clamps (green and orange). The end without non-return valve allows the sampling blood.
Length of tubes: 6 cm / 6 cm
Used for peripheral cannula care, Octopus 2 has the following advantages:
- Eliminates the risk of access-associated needle-stick injury
- Reduces the risk of cannula movement in the vein causing mechanical vein irritation
- Negates the need to use the cannula injection port (non disinfectable), Bionector is disinfectable
- Eliminates the need to use a 3-way tap to run simultaneous compatible solutions
- Provides closed system (avoiding blood reflux) when accessing the catheter
- PUR material (no PVC, DEHP-free - no release of softeners)
Non-return valves avoid reflux of fluids or drugs beyond the valve during multiple infusions.
Bionector can be disinfected with most of the antiseptic solutions (example: 70% alcohol). Avoid using antiseptic containing both alcohol and hydrogen peroxide.
For more information about Bionector, please refer to technical file code 896.
Octopus with Bionector are also available with 1 lumen (code 5222.014), 3 lumen (codes 841.364-842.311/312), 4 lumen (code 842.414) and 5 lumen (code 842.514).
Code |
Extension tube
Total priming volume ml |
Lumen 1
Flowrate ml/min |
Lumen 1
Length cm |
Lumen 1
Int. Ø mm |
Lumen 1
External diameter mm |
Lumen 1
Priming volume ml |
Lumen 2
Flowrate ml/min |
Lumen 2
Length cm |
Lumen 2
Int. Ø mm |
Lumen 2
External diameter mm |
Lumen 2
Priming volume ml |
---|---|---|---|---|---|---|---|---|---|---|---|
841.232 | 0.50 | 78 | 3 | 1.50 | 2.50 | 0.30 | 78 | 3 | 1.50 | 2.50 | 0.30 |
841.264 | 0.57 | 82 | 10 | 1.50 | 2.50 | 0.34 | 82 | 10 | 1.50 | 2.50 | 0.34 |
842.209 | 0.58 | 73 | 6 | 1.50 | 2.50 | 0.36 | 78 | 6 | 1.50 | 2.50 | 0.33 |
Extension tube | Lumen 1 | Lumen 2 | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Code | Total priming volume ml |
Flowrate ml/min |
Length cm |
Int. Ø mm |
External diameter mm |
Priming volume ml |
Flowrate ml/min |
Length cm |
Int. Ø mm |
External diameter mm |
Priming volume ml |
|
841.232 | 0.50 | 78 | 3 | 1.50 | 2.50 | 0.30 | 78 | 3 | 1.50 | 2.50 | 0.30 | |
841.264 | 0.57 | 82 | 10 | 1.50 | 2.50 | 0.34 | 82 | 10 | 1.50 | 2.50 | 0.34 | |
842.209 | 0.58 | 73 | 6 | 1.50 | 2.50 | 0.36 | 78 | 6 | 1.50 | 2.50 | 0.33 |
- Contains Latex: No
- Contains DEHP: No
- Contains biogical or animal-based product: No
- Non-pyrogenic: Yes
- 385.54 Ko Brochure Octopus Neonat